xAM-imaging

Repository containing script that implements the xAM imaging sequence for artifact-free detection of gas vesicles (GVs). By transmitting cross-propagating plane waves, xAM suppresses nonlinear propagation artifacts while preserving nonlinear contrast from GVs.

License: MIT Made with MATLAB Vantage

Prerequisites

Make sure you have the following software installed.

  • MATLAB: R2021b or later
  • CUDA Toolkit: Version ≥ 12.0 (recommended: CUDA 12.9.1)
  • Verasonics Vantage: Version 4.0 or later

How to use

1. Parameter Review

Open the imaging sequence xAM_imaging.m and review the following parameters:

Trans.name = 'L22-14vX';        % tested transducers are L22-14v / L11-5v
P.image_start_depth_mm = 0;     % start depth [mm]
P.image_end_depth_mm = 10;      % end depth [mm]
P.xwave_angles = 17;            % xWave angle in degrees, check if max depth can be reached! 
P.speed_of_sound = 1480;        % agar/water 1480 m/s, tissue 1540 m/s
P.image_voltage = 2.5;          % set to safe number to avoid collapse
P.save_path = 'data';           % default path for data saving

Other parameters that can be configured are described below.

2. Launch imaging sequence

Run the script xAM_imaging.m to start imaging. The VSX GUI will pop up.

VSX gui

When it is safe, if the ultrasound transducer is in contact with the phantom, water or other, press the button "Start Sequence". Next, xAM data will be acquired continuously and displayed on screen, side by side with a B-mode image.

GUI Controls

Control Description
Start Seq. Starts the imaging sequence. After launching the script, the GUI window opens (if parameters are valid). When the transducer is in contact with the imaging medium, press Start Seq. to unpause the hardware and begin continuous acquisition.
Freeze/Unfreeze Pauses or resumes the sequence. Only works after the sequence has been started with Start Seq.
Save Data Opens the save menu to set path and filename. Requires the sequence to be frozen before saving; the current on-screen image will then be saved.
Imaging Voltage Controls the transmit voltage. Start with a low voltage to avoid collapsing gas vesicles. At too low voltages, nonlinear signal in the xAM images may not appear since the GV buckling threshold is not reached. Slowly increase the voltage until nonlinear signal is observed.
Persistence Controls image persistence by blending new frames with previous ones. 1.0 = only new frame (no persistence), 0.0 = only old frame. Intermediate values apply a moving average for smoother, more stable images.
Dynamic Range B-mode Adjusts the dynamic range used for display of the B-mode image.
Dynamic Range xAM Adjusts the dynamic range used for display of the xAM image.

Parameter Reference

Name Description
image_start_depth_mm Start depth of imaging region [mm]
image_end_depth_mm End depth of imaging region [mm]
xwave_angles xWave angle [degrees]
speed_of_sound Speed of sound in medium (1480 m/s in water/agar, 1540 m/s in tissue)
image_voltage Transmit voltage [V]. Set safely to avoid GV collapse
aperture_size_min Minimum number of elements in active aperture (e.g. 48 for wide FOV)
aperture_size_max Maximum number of elements in active aperture
transmit_apodization Apodization function to reduce edge waves. Options:
none, kaiser, hamming, tukey
fps Acquisition frame rate [Hz]
save_path Default path for saving data

Advanced options

Name Description
num_accumulations Number of RF accumulations. Increases SNR but lowers FPS and risks clipping of signals
use_adaptive_xwave_angle If true, xWave angles are adapted based on imaging depth
use_half_pitch_scanning Enables half-pitch scanning for finer sampling
transmit_frequency Transducer transmit frequency [MHz]

Notes

  • Tested on MATLAB R2021b on Windows 10, with Verasonics Vantage-4.8.6.
  • You require a Verasonics host PC with GPU, since the sequence uses a GPU accelerated beamformer.

Feedback, Feature Requests, and Bug Reports

We welcome contributions from users and developers!

  • Found a bug? Please open an issue with a clear description and steps to reproduce.
  • Have a feature request? Open an issue and label it as enhancement.
  • General feedback or suggestions? Use the feedback template to share your thoughts.

Your input helps improve this repository and guide future development.

References to be cited

  • Maresca, D., Sawyer, D. P., Renaud, G., Lee-Gosselin, A., & Shapiro, M. G. (2018). Nonlinear X-wave ultrasound imaging of acoustic biomolecules. Physical Review X, 8(4), 041002, DOI: 10.1103/PhysRevX.8.041002
  • Matalliotakis, A., Waasdorp, R., Verweij, M. D., & Maresca, D. (2024). Impact of wavefront shape on nonlinear ultrasound imaging of monodisperse microbubbles. Physical Review Applied, 22(3), 034062, DOI: 10.1103/PhysRevX.8.041002

Disclaimer

This software is provided by the authors and contributors "as is" and any express or implied warranties, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose are disclaimed. In no event shall the authors and contributors be liable for any direct, indirect, incidental, special, exemplary, or consequential damages (including, but not limited to, procurement of substitute goods or services; loss of use, data, or profits; or business interruption) however caused and on any theory of liability, whether in contract, strict liability, or tort (including negligence or otherwise) arising in any way out of the use of this software, even if advised of the possibility of such damage.

About

Implementation of cross-amplitude modulation (xAM) ultrasound imaging for artifact-free visualization of gas vesicles (GVs).